The Ivoclar Programat P90 Go to product viewer dialog for this item.
Use the green START key to begin a program and the red STOP key to abort it or silence the buzzer.
The P90 uses a flashing display to indicate specific functional failures. Error Code Likely Fix T-value is lower than actual temperature Adjust the program's holding temperature. Er03 Vacuum switch-off value exceeds temperature Ensure vacuum settings do not exceed preset temperature T. Er06 B-value (standby) is out of range Enter a standby temperature between 100°C and 700°C. Er12 Low battery in control unit Replace the internal battery immediately. Er13 Excessive temperature detected Check for programming errors or cooling system blockages. Operational Quick Guide ivoclar programat p90 manual fixed
Turn the furnace on and wait for the buzzer to sound before placing ceramics on the firing mount.
If the P90 displays Error Er12 , the internal battery in the control unit is low and must be replaced. The Ivoclar Programat P90 Go to product viewer
If the vacuum cannot be established, check the sealing rim (69) and sealing ring (35) for debris or damage. Clean these components regularly with a dry or slightly moist cloth; never use solvents. Troubleshooting Common Error Codes
The heating muffle should be replaced if heating becomes inadequate or if the furnace displays Er23 or Er24 . Avoid touching the quartz tube or heating wire with bare fingers to prevent premature damage from grease contamination. Error Code Likely Fix T-value is lower than
Maintaining the P90 involves routine checks and occasional hardware replacements to prevent firing errors.